8000Q2

Availability: In Stock
Reusable Ear Clip Pulse Oximetry Sensor

The Nonin 8000Q2 reusable ear clip sensor offers flexibility for spot check and longer-term SpO2 monitoring when digits are not the preferred application site. With Nonin PureLight® sensor technology and PureSAT® signal processing technology, you can be sure you’re getting consistent results from patient to patient and sensor to sensor.

Compare
Category:

SpO2 Accuracy:1, 2, 3

Range Oxygen Saturation (Arms*)
70 – 100% ±3
70 – 80% ±3
80 – 90% ±4
90 – 100% ±3

SpO2 Low Perfusion Accuracy: 70% to 100% ±2 digits (Arms*)1
Pulse Rate Accuracy: 40 BPM to 240 BPM ±3 digits (Arms*)1
Pulse Rate Low Perfusion Accuracy: 40 BPM to 240 BPM ±3 digits (Arms*)1
Temperature: 4, 5
Operating: -20 °C to 50 °C (-4 °F to 122 °F)
Storage/Transportation: -40 °C to 70 °C (-40 °F to 158 °F)

Humidity: 4, 5
Operating: 10% to 95% non-condensing
Storage/Transportation: 10% to 95% non-condensing

* ±1 Arms encompasses 68% of the population at zero bias.
1 SpO2 accuracy was conducted during induced hypoxia study on healthy subjects over the range of 70 % to 100 %.
2 Accuracy specifications based on Nonin’s PureSAT® SpO2 technology and PureLight® sensor technology.
3 Additional accuracy and performance information can be found in the sensor accuracy document on the operator’s manual CD.
4 For combined oximeter/sensor specifications, refer to the applicable oximetry system’s operator’s manual.
5 Range as tested with Nonin’s PureSAT SpO2 technology.

Measurement Wavelengths and Output Power**:
Red: 660 nanometers @ 0.8 mW nominal
Infrared: 910 nanometers @ 1.2 mW nominal
** This information is especially useful for clinicians performing photodynamic therapy.

Compliance
This product complies with ISO 10993-1.
Not made from natural rubber latex.

Documentation

Videos

8000Q2 Ear Clip Pulse Oximeter Sensor

Resources