Oxygen Saturation Display Range: 0 to 100% SpO2
Pulse Rate Display Range: 18 to 321 pulses per minute (BPM)
Displays:
Pulse Quality: LED, amber
Sensor Alarm: LED, amber
Pulse Strength Bargraph: LED, bargraph, tri-color segments
Alarm Indicator: LED, bi-color
Alarm Silenced: LED, amber
Numeric Displays: 3-digit, 7-segment LEDs, green
Low Battery: LED, amber
Accuracy – Sensors: Declared accuracy for compatible sensors can be foundin Nonin’s Sensor Accuracy document.
Alarm Volume:
High: 75 dBA
Low: 63 dBA
Informational Tone Volume:
High: 67 dBA
Low: 55 dBA
Measurement Wavelengths and Output Power:a
Red: 660 nm @ 0.8 mW maximum average power
Infrared: 910 nm @ 1.2 mW maximum average power
Memory: 70 hours (assuming continuous operation)
Temperature:
Operating: 0 °C to 40 °C (32 °F to 104 °F)
Storage/Transportation: -40 °C to 70 °C (-40 °F to 158 °F)
Time (from storage) for monitor to be ready for its intended use:
6 minutes to warm from -40 °C to 0 °C
23 minutes to cool from 70 °C to 40 °C
Humidity:
Operating: 10 to 90% noncondensing
Storage/Transportation: 10 to 95% noncondensing
Altitude (Operating): Up to 4,000 meters (13,123 feet)
Hyperbaric Pressure: Up to 4 atmospheres
a. This information is especially useful for clinicians performing photodynamic therapy.
Power Requirements (Mains): 100-240 VAC 50-60 Hz
Internal Power:
Battery: 7.2 volt NiMH battery pack
Operating Life (fully charged battery): 16 hours minimum
Storage Life: 21 days minimum
Recharge Rate: 4 hours maximum
Dimensions: Approximately 219 mm (8.6”) W x 92 mm (3.6”) H x 142 mm (5.6”) D
Weight: Approximately 900 grams (2 lbs) with battery
Warranty: 3 years
Classification per ANSI/AAMI ES60601-1 and CAN/CSA-C22.2 No. 60601-1:
Type of Protection: Class II (on AC power with MPP30 power supply). Internally powered (on battery power)
Degree of Protection: Type BF-Applied Part
Mode of Operation: Continuous
Enclosure Degree of Ingress Protection: IP33
Analog Outputs:
SpO2 Output Range: 0-1 VDC (0-100% SpO2), 1.27 VDC (out of track)
Pulse Rate Output Range: 0-1 VDC (0-300 BPM), 1.27 VDC (out of track)
Event Maker: 0 V (no event), 1 V (event occurred)
Accuracy: ±2% (SpO2), ±5% (Pulse Rate)
Load Current: 2 mA maximum