SpO2 Accuracy*:
The table below shows Arms values measured using the Model 8100S(X) in a clinical study.
Range | Oxygen Saturation (Arms**) | Motion Oxygen Saturation (Arms**) | Low Perfusion Oxygen Saturation(Arms**) |
---|---|---|---|
70 – 100% | ±2 | ±3 | ±2 |
70 – 80% | ±2 | ±3 | ±2 |
80 – 90% | ±2 | ±3 | ±2 |
90 – 100% | ±2 | ±2 | ±2 |
Pulse Rate Accuracy*:
Non-motion: 18 – 300 BPM ±3 digits (Arms**)
Motion: 40 – 240 BPM ±5 digits (Arms**)
Low perfusion: 40 – 240 BPM ±3 digits (Arms**)
Temperature:
Operating: 0 °C to 40 °C (32 °F to 104 °F)
Storage/Transportation: -30 °C to 70 °C (-22 °F to 158 °F)
Humidity:
Operating: 15 – 93% non-condensing
Storage/Transportation: Up to 93% non-condensing
Operating Altitude: 0 to 4,000 m (0 to 13,123 ft)
* Additional accuracy and performance information can be found in the system operator’s manual.
** ±1 Arms encompasses 68% of the population.
Measurement Wavelengths and Output Power***:
Red: 660 nanometers @ 0.8 mW nominal
Infrared: 910 nanometers @ 1.2 mW nominal
*** This information is especially useful for clinicians performing photodynamic therapy.
Compliance
This product complies with ISO 10993-1.
Not made from natural rubber latex.